首页   按字顺浏览 期刊浏览 卷期浏览 Spark-Source Mass Spectrometric Techniques for the Analysis of Lanthanide and Actinide ...
Spark-Source Mass Spectrometric Techniques for the Analysis of Lanthanide and Actinide Elements in Microgram and Sub-Microgram Transuranium Samples

 

作者: JoelA. Carter,   A.E. Cameron,   JohnA. Dean,  

 

期刊: Analytical Letters  (Taylor Available online 1971)
卷期: Volume 4, issue 6  

页码: 337-350

 

ISSN:0003-2719

 

年代: 1971

 

DOI:10.1080/00032717108058621

 

出版商: Taylor & Francis Group

 

关键词: mass spectroscopy;spark source;analysis;actinides;lanthanides;transuranium;solutions

 

数据来源: Taylor

 

摘要:

Lanthanide and actinide impurities in a total sample of 1 to 10 μg of transuranic matrix can be determined by employing the spark-source mass spectrograph and solution technique. A special source glove box permits safe manipulation of transuranium elements with α-emitting radioactivity up to 1012dpm. The minimum detection limit is about 1012atoms per electrode system and an exposure of 10−8coulomb. Relative sensitivity values for the lanthanides and actinides were established with respect to erbium, the internal standard.

 