Spark-Source Mass Spectrometric Techniques for the Analysis of Lanthanide and Actinide Elements in Microgram and Sub-Microgram Transuranium Samples
作者:
JoelA. Carter,
A.E. Cameron,
JohnA. Dean,
期刊:
Analytical Letters
(Taylor Available online 1971)
卷期:
Volume 4,
issue 6
页码: 337-350
ISSN:0003-2719
年代: 1971
DOI:10.1080/00032717108058621
出版商: Taylor & Francis Group
关键词: mass spectroscopy;spark source;analysis;actinides;lanthanides;transuranium;solutions
数据来源: Taylor
摘要:
Lanthanide and actinide impurities in a total sample of 1 to 10 μg of transuranic matrix can be determined by employing the spark-source mass spectrograph and solution technique. A special source glove box permits safe manipulation of transuranium elements with α-emitting radioactivity up to 1012dpm. The minimum detection limit is about 1012atoms per electrode system and an exposure of 10−8coulomb. Relative sensitivity values for the lanthanides and actinides were established with respect to erbium, the internal standard.
点击下载:
PDF (1051KB)
返 回