首页   按字顺浏览 期刊浏览 卷期浏览 Digital frequency tripling circuit for third harmonic detection by lock‐in ampli...
Digital frequency tripling circuit for third harmonic detection by lock‐in amplifiers

 

作者: Seung‐Min Lee,   Sook‐Il Kwun,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1994)
卷期: Volume 65, issue 4  

页码: 971-973

 

ISSN:0034-6748

 

年代: 1994

 

DOI:10.1063/1.1144928

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A new digital circuit for fast and stable frequency tripling was constructed. The frequency tripler covers the input frequency range from 0.1 Hz up to 5 kHz with a maximum phase shift less than 1 deg at the highest frequency. The tripler generates a stable signal synchronized to the input after just one input period. Since the circuit design is flexible, input frequency range and phase accuracy can be improved using faster integrated circuits.  

 