Cotside plasma rather than whole blood estimates of blood glucose in the newborn
作者:
G. J. REYNOLDS,
G. KIAGI,
期刊:
Journal of Paediatrics and Child Health
(WILEY Available online 1994)
卷期:
Volume 30,
issue 1
页码: 33-35
ISSN:1034-4810
年代: 1994
DOI:10.1111/j.1440-1754.1994.tb00562.x
出版商: Blackwell Publishing Ltd
关键词: blood glucose;hypoglycaemia;infant;newborn
数据来源: WILEY
摘要:
AbstractAn evaluation of the use of a glucose reflectance meter for the cotside measurement of plasma rather than whole blood glucose was undertaken. Three hundred and twelve samples were obtained for examination. There is a closer correlation between plasma glucose than whole blood glucose to laboratory values. Confidence and prediction intervals show that even separating plasma from whole blood at the cotside does not improve the reliability of glucose reflectance meters when predicting blood glucose values in the context of managing high risk newborn infants. The cotside measurement of blood glucose using glucose reflectance meters should be abandoned and improved methods for quickly obtaining accurate laboratory data should be instituted.