首页   按字顺浏览 期刊浏览 卷期浏览 Profile of a focussed collimated laser beam near the focal minimum characterized by flu...
Profile of a focussed collimated laser beam near the focal minimum characterized by fluorescence correlation spectroscopy

 

作者: Stanley M. Sorscher,   Melvin P. Klein,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1980)
卷期: Volume 51, issue 1  

页码: 98-102

 

ISSN:0034-6748

 

年代: 1980

 

DOI:10.1063/1.1136028

 

出版商: AIP

 

数据来源: AIP

 

摘要:

Central to the application of fluorescence correlation spectroscopy, to measure the self‐diffusion coefficients and average concentration of fluorescent molecules in a volume determined by a focussed laser beam, is the determination of the focal spot size. As the focal spot size in the sample plane is varied by displacing either the focusing lens or sample position along the beam axis, the diffusion time and average number of molecules vary in a parabolic manner. Analysis of the parameters of the parabola leads to estimates of the beam radius at the waist. The results agree with theoretical predictions and provide an independent measurement of the beam profile.

 

点击下载:  PDF (306KB)



返 回