An Absolute Method of Measuring Energy Outputs from CO2Lasers
作者:
J. H. Jacob,
E. R. Pugh,
J. D. Daugherty,
D. B. Northam,
期刊:
Review of Scientific Instruments
(AIP Available online 1973)
卷期:
Volume 44,
issue 4
页码: 471-474
ISSN:0034-6748
年代: 1973
DOI:10.1063/1.1686158
出版商: AIP
数据来源: AIP
摘要:
In the energy meter described here the CO2laser radiation is almost completely absorbed by an Al2O3surface on an aluminum plate. The high absorption of the surface (≥95%) eliminates the need for calibration and the laser energy is computed by measuring the temperature rise of a plate with known thermal capacity. Experimentally we find that the calorimeter noise level corresponds to an energy density of about 2 mJ/cm2. Calculations indicate that the instrument is capable of measuring the energy of a single pulse that is 1 nsec wide, provided the pulse contains an average energy density of 10 mJ/cm2. We have estimated the radiation and convection losses and found them to be negligible. The calorimeter is good for radiation between 8.5 and 11 &mgr;; however, with modifications the bandwidth could easily be increased.
点击下载:
PDF
(274KB)
返 回