首页   按字顺浏览 期刊浏览 卷期浏览 An Absolute Method of Measuring Energy Outputs from CO2Lasers
An Absolute Method of Measuring Energy Outputs from CO2Lasers

 

作者: J. H. Jacob,   E. R. Pugh,   J. D. Daugherty,   D. B. Northam,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1973)
卷期: Volume 44, issue 4  

页码: 471-474

 

ISSN:0034-6748

 

年代: 1973

 

DOI:10.1063/1.1686158

 

出版商: AIP

 

数据来源: AIP

 

摘要:

In the energy meter described here the CO2laser radiation is almost completely absorbed by an Al2O3surface on an aluminum plate. The high absorption of the surface (≥95%) eliminates the need for calibration and the laser energy is computed by measuring the temperature rise of a plate with known thermal capacity. Experimentally we find that the calorimeter noise level corresponds to an energy density of about 2 mJ/cm2. Calculations indicate that the instrument is capable of measuring the energy of a single pulse that is 1 nsec wide, provided the pulse contains an average energy density of 10 mJ/cm2. We have estimated the radiation and convection losses and found them to be negligible. The calorimeter is good for radiation between 8.5 and 11 &mgr;; however, with modifications the bandwidth could easily be increased.

 

点击下载:  PDF (274KB)



返 回