首页   按字顺浏览 期刊浏览 卷期浏览 Perchlorate Ion-Selective Electrodes with the Liquid Membranes of the O-Phenanthroline ...
Perchlorate Ion-Selective Electrodes with the Liquid Membranes of the O-Phenanthroline Chelate or its Related Compounds

 

作者: Nobuhiko Ishibashi,   Hitoshi Kohara,  

 

期刊: Analytical Letters  (Taylor Available online 1971)
卷期: Volume 4, issue 11  

页码: 785-792

 

ISSN:0003-2719

 

年代: 1971

 

DOI:10.1080/00032717108058678

 

出版商: Taylor & Francis Group

 

关键词: electrode;perchlorate;o-phenanthroline;bathophenanthroline

 

数据来源: Taylor

 

摘要:

The metal chelates of o-phenanthroline, a, a′-dipyridyl or bathophenanthroline were used as the ion exchanger in the liquid membrane of the perchlorate ion-selective electrode. The electrode with a nitrobenzene membrane containing tris(bathophenanthroline)ferrous perchlorate is the highest sensitive one and gives a linear Nernstian response up to about 1 × 10−5Mperchlorate. The membrane electrode having the ferrous ion-chelate of o-phenanthroline as an ion exchanger shows an excellent selectivity for perchlorate ion over nitrate or iodide Ion. The effects of the chelate concentration in the membrane and the central metal species of the chelate are examined on the electrode performance.

 

点击下载:  PDF (236KB)



返 回