Assessment of Hitachi HME-20 pulse and blood-pressure monitor
作者:
NorthcoteRobin J.,
O'donoghueJoseph,
BallantyneDavid,
期刊:
Journal of Medical Engineering&Technology
(Taylor Available online 1985)
卷期:
Volume 9,
issue 1
页码: 10-13
ISSN:0309-1902
年代: 1985
DOI:10.3109/03091908509018200
出版商: Taylor&Francis
关键词: Blood Pressure;Hitachi HME-20 Monitor
数据来源: Taylor
摘要:
The performance of the Hitachi HME-20 pulse and blood-pressure (BP) monitor in comparison with direct intra-arterial BP recording and electrocardiographic monitoring is described. Highly significant (p<0·001) correlations were found between intra-arterial systolic and diastolic pressures and pressures recorded by the Hitachi monitor. Similarly, the electrocardiographically computed heart-rate, and that given by the Hitachi monitor were significantly correlated (p150 mmHg. These results are comparable to more expensive pulse and blood-pressure monitors. We conclude that the instrument can reproduce a satisfactory estimate of heart-rate and blood-pressure and may be of particular use when a change of blood-pressure is of prime importance, rather than an absolute measurement.