Correction of fluorescence detected circular dichroic photoselection artifacts using light collecting ellipsoidal mirrors

 

作者: S. E. Bicknese,   M. F. Maestre,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1987)
卷期: Volume 58, issue 11  

页码: 2060-2063

 

ISSN:0034-6748

 

年代: 1987

 

DOI:10.1063/1.1139463

 

出版商: AIP

 

数据来源: AIP

 

摘要:

Ellipsoidal mirrors are used to collect the emitted light from fluorophores in fluorescence detected circular dichroic (FDCD) measurements and to deliver this light to a photomultiplier tube (PMT) at 90° to the excitation beam. These mirrors collect light over a symmetrical 71% of the emission sphere centered on the sample cell, and thereby eliminate most of the nonuniformity in detected emission resulting from the anisotropy of photoselection. Furthermore, even with reflectance and path length losses, the amount of fluorescent emission reaching the PMT is increased by a factor of approximately 18. Elimination of photoselection artifacts in UV FDCD spectra is demonstrated by comparing spectra of L‐tryptophan in glycerol with its spectra in water. Validity of the values measured is verified through calibration with a d10 camphorsulfonic acid (d10 CSA) standard.

 