Calibration Designs for Machines with Carry-Over and Drift
作者:
DanielCuthbert,
期刊:
Journal of Quality Technology
(Taylor Available online 1975)
卷期:
Volume 7,
issue 3
页码: 103-108
ISSN:0022-4065
年代: 1975
DOI:10.1080/00224065.1975.11980679
出版商: Taylor&Francis
关键词: Calibration Designs;Sequenced Calibration
数据来源: Taylor
摘要:
Continuous-flow analysers (like the Auto Analyser and its current version SMAC) require individual calibration to verify the shape of their calibration curves, of the amount of sample-to-sample carryover, and of the extent of linear and quadratic time-drift within each sequence of (46) analyses. A sequence of 46 trials, using five calibrating concentrations is given which permits simultaneous estimates of all these attributes of the system.
点击下载:
PDF (709KB)
返 回