Ion-Selective Electrodes for the Determination of Antimony and Thallium Based on Water-Insoluble Basic Dye Salts
作者:
A.G. Fogg,
A.A. Al-sibaai,
C. Burgess,
期刊:
Analytical Letters
(Taylor Available online 1975)
卷期:
Volume 8,
issue 2
页码: 129-137
ISSN:0003-2719
年代: 1975
DOI:10.1080/00032717508058471
出版商: Taylor & Francis Group
关键词: ion-selective electrodes;antimony;thallium;basic dye
数据来源: Taylor
摘要:
Hexachloroantimonate(V) and tetrachlorothallate(III) salts of Sevron Red L (C.I. Basic Red 17), Sevron Red GL (C.I. 11,085), Flavinduline O (C.I. 50,000) and Phenazinduline O have been applied in liquid-state ion-selective electrodes for the determination of antimony and thallium. The slope of potential versus pM graphs approach full Nernstian response in the range 10 M (Sb) and 10−6M (T1) to 10−2M. An unusual feature of the electrodes is their response to antimony(III) as well as to antimony(V).