首页   按字顺浏览 期刊浏览 卷期浏览 Pulse oximetry during low cardiac output and hypothermia states immediately after open ...
Pulse oximetry during low cardiac output and hypothermia states immediately after open heart surgery

 

作者: HEIRKI PALVE,   ARNO VUORI,  

 

期刊: Critical Care Medicine  (OVID Available online 1989)
卷期: Volume 17, issue 1  

页码: 36-69

 

ISSN:0090-3493

 

年代: 1989

 

出版商: OVID

 

数据来源: OVID

 

摘要:

Two commercially available pulse oximeters were found to read reliably early during periods of low cardiac index (CI) and low peripheral temperature (Tp). The lowest mean CI and mean Tp values at which a reading could be obtained were 2.4 L/min m: and 26.5°C. respectively. The highest mean systemic vascular resistance index was 2930 dyne -sec/cm5m2. The reliability of one monitor was equally good both in a normal and low perfusion mode (i.e. when the monitor displayed a low-quality signal message). The use of locally applied vasodilating creme widened the physiologic limits within which reliable measurements could be obtained. All oximeters and sensors overestimated slightly the saturation as compared to the hemoximeter readings. possibly because of a high mean carboxyhemoglobin concentration that was found to be present.

 

点击下载:  PDF (260KB)



返 回