Low temperature thermometry in high magnetic fields. IV. Allen‐Bradley carbon resistors (0.5–4.2 K)
作者:
H. H. Sample,
L. J. Neuringer,
期刊:
Review of Scientific Instruments
(AIP Available online 1974)
卷期:
Volume 45,
issue 11
页码: 1389-1391
ISSN:0034-6748
年代: 1974
DOI:10.1063/1.1686509
出版商: AIP
数据来源: AIP
摘要:
The transverse magnetoresistance of six 1/8‐W Allen‐Bradley carbon resistors with nominal room temperature resistances of 2.7, 3.9, and 5.6 &OHgr; was measured at various temperatures between 0.5 and 4.2 K in static magnetic fields up to 140 kG. Below 1.5 K, the magnetic field induced temperature errors of the resistors are small, less than 25%, and are very reproducible, i.e., ±0.8% in the case of the 2.7 &OHgr; units. Hence, these resistors are useful as low temperature thermometers in strong magnetic fields.
点击下载:
PDF
(135KB)
返 回