首页   按字顺浏览 期刊浏览 卷期浏览 Enzyme Electrode Sensor for Hydrogen Peroxide
Enzyme Electrode Sensor for Hydrogen Peroxide

 

作者: A. Ciucu,   V. Magearu,   C. Luca,  

 

期刊: Analytical Letters  (Taylor Available online 1985)
卷期: Volume 18, issue 3  

页码: 299-313

 

ISSN:0003-2719

 

年代: 1985

 

DOI:10.1080/00032718508066133

 

出版商: Taylor & Francis Group

 

关键词: enzymatic sensor;2,4-dichloro-6-methoxy-s-triazine;hydrogen peroxide;rapid;amperometric

 

数据来源: Taylor

 

摘要:

An enzymatic sensor was constructed from an oxygen amperometric sensor on whose surface a dialysis membrane containing covalently bonded catalase was set. Immobilization of the enzyme on the dialysis membrane surface was achieved with 2,4-dichloro-6-methoxy-s-triazine, a derivative of cyanuric chloride, which unlike others of its monosubstituted derivatives ensures some advantages. The time of measurement is less than 12 sec, for the kinetic method of the initial slope and one minute for the steady-state method. The sensor responds linearly to hydrogen peroxide in the concentration range 10−3- 10−5moles/1. The utilization of this sensor in intermittent and continuous operation in a laboratory enzymatic minireactor is discussed.

 

点击下载:  PDF (389KB)



返 回