Central Corneal Thickness Measurement with a Retinal Optical Coherence Tomography Device Versus Standard Ultrasonic Pachymetry

 

作者: Martin Bechmann,   Martin Thiel,   Aljoscha Neubauer,   Stephanie Ullrich,   Klaus Ludwig,   Kenneth Kenyon,   Michael Ulbig,  

 

期刊: Cornea  (OVID Available online 2001)
卷期: Volume 20, issue 1  

页码: 50-54

 

ISSN:0277-3740

 

年代: 2001

 

出版商: OVID

 

关键词: Optical coherence tomography;Central corneal thickness;Ultrasonic pachymetry;Corneal edema

 

数据来源: OVID

 

摘要:

Purpose.To demonstrate the capability of a standard, commercially available optical coherence tomography device (originally designed to measure retinal thickness) to image the human cornea in vivo and to measure central corneal thickness (CCT) in normal and edematous corneas. The intrapatient precision and interpatient variability of this novel application was compared to standard ultrasonic pachymetry. Also, the correlation of both methods was investigated.Methods.CCT measurements using optical coherence tomography (OCT) and ultrasonic pachymetry (PACH) were obtained in 36 normal eyes and 16 eyes with corneal edema.Results.Direct in vivo imaging of the cornea and measurements of CCT by OCT could be performed in all eyes. In normal subjects, CCTOCTwas 530 ± 32 &mgr;m and CCTPACHwas 581 ± 34 &mgr;m. The two methods showed a highly significant correlation with a standardized regression coefficient of 0.988. The difference between both methods was constant over the range of CCT (mean difference, 49.4 ± 5.9 &mgr;m). The mean intrapatient SD of CCT measurements was 4.90 &mgr;m and 4.96 &mgr;m for OCT and PACH, respectively. In patients with corneal edema, mean CCTOCTwas 601 ± 59 &mgr;m, and mean CCTPACHwas 667 ± 68 &mgr;m. The difference between the two techniques increased slightly with increasing corneal edema (mean difference, 66.9 ± 10.9 &mgr;m).Conclusion.Imaging of the human cornea can be performed by a standard retinal OCT device, and OCT measurement of CCT shows excellent correlation to values obtained by PACH, giving similar readings separated by a constant difference. In corneal edema, the difference between the two methods is additionally increased, but continues to demonstrate excellent consistency.

 

点击下载:  PDF (818KB)



返 回