The Measurement of Extremely Low Light Levels in Analytical Chemi- and Bio-Luminescence: A Mathematical and Practical Evaluation of Some Physical Factors Influencing Detector Output and Instrument Design
作者:
D.G. Bullock,
R.A. Bunce,
T.J. N. Carter,
期刊:
Analytical Letters
(Taylor Available online 1979)
卷期:
Volume 12,
issue 8
页码: 841-854
ISSN:0003-2719
年代: 1979
DOI:10.1080/00032717908059765
出版商: Taylor & Francis Group
关键词: Chemiluminescence;bioluminescence;instrument design;luminometer;mathematical modelling
数据来源: Taylor
摘要:
In order to study the design features which affect the sensitivity of chemi- and bio-luminescence measuring devices, mathematical models have been devised. These models have been tested experimentally, and good agreement with their predictions obtained. Features such as dimensions of source and detector, distance from source to detector, and the use of mirrors have been investigated. This approach has provided a rational basis for the design of luminescence measuring devices.