首页   按字顺浏览 期刊浏览 卷期浏览 An isothermal scanning calorimeter controlled by linear pressure variations from 0.1 to...
An isothermal scanning calorimeter controlled by linear pressure variations from 0.1 to 400 MPa. Calibration and comparison with the piezothermal technique

 

作者: Stanislaw L. Randzio,   Jean‐Pierre E. Grolier,   Jacques R. Quint,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1994)
卷期: Volume 65, issue 4  

页码: 960-965

 

ISSN:0034-6748

 

年代: 1994

 

DOI:10.1063/1.1144926

 

出版商: AIP

 

数据来源: AIP

 

摘要:

An isothermal scanning calorimeter controlled by linear pressure variations is described for the pressure range 0.1–400 MPa at temperatures from 303 to 573 K. The rate of pressure variations can be as low as 0.002 MPa/s over the whole pressure range. The functioning of the instrument was tested by measuring the coefficient of thermal expansivity of liquidn‐hexane, with calibration performed with gaseous nitrogen and by measuring the coefficient of thermal expansivity of benzene with calibration performed with liquidn‐hexane. The results are compared with literature results obtained by the piezothermal technique and with results obtained in the present instrument by the use of a pressure stepwise scanning mode. An example of an investigation of an isothermal solid‐to‐liquid transition in benzene is also given.

 

点击下载:  PDF (616KB)



返 回