Traditional Liquid Quality Control Versus Internal Quality ControlA Clinical Evaluation of the Radiometer NPT7

 

作者: James Nichols,   Karen Dyer,   Sandra Humbertson,   Juanita Stem,   Thomas Krarup,   Peter Frischauf,  

 

期刊: Point of Care: The Journal of Near-Patient Testing & Technology  (OVID Available online 2002)
卷期: Volume 1, issue 1  

页码: 9-19

 

ISSN:1533-029X

 

年代: 2002

 

出版商: OVID

 

关键词: Blood gas;Critical care;Quality control;Point of Care Testing.

 

数据来源: OVID

 

摘要:

&NA;The NPT7 (Radiometer Medical A/S, Copenhagen, Denmark) point of care blood gas analyzer was evaluated in a tertiary care academic institution. The NPT7 demonstrated comparable precision to the ABL720 analyzers (coefficients of variation (CV) range = 0 to 4.9% and 0 to 7.8% for the NPT7 and ABL720 respectively). Correlation was excellent with regression slopes (0.88 to 1.05), intercepts (−2.76 to +7.60) and correlation coefficients (r = 0.9287 to 0.9980) (CV for duplicate differences range = 0.07 to 6.3% and 0.04 to 5.0% for the NPT7 and ABL720 analyzers excluding carboxyhemoglobin, methemoglobin, and deoxyhemoglobin analytes which were primarily in a low normal range). The internal QualityGuard system detected more errors (average 22 failures per analyzer) than traditional liquid controls (3 events). QualityGuard was able to monitor the entire analytical process and substitute for liquid controls in guaranteeing the quality of test results. With no required maintenance and intuitive operator interface, the NPT7 has application in a variety of settings.

 

点击下载:  PDF (1614KB)