Response of a chevron microchannel plate to 2.5 and 14 MeV neutrons
作者:
S. S. Medley,
R. Persing,
期刊:
Review of Scientific Instruments
(AIP Available online 1981)
卷期:
Volume 52,
issue 10
页码: 1463-1468
ISSN:0034-6748
年代: 1981
DOI:10.1063/1.1136476
出版商: AIP
数据来源: AIP
摘要:
The response of a large area (4.6×13 cm) multianode channel electron multiplier array (CEMA) detector to energetic neutrons was investigated. The measured neutron detection efficiencies of the chevron microchannel plate (MCP) were 1.7×10−3and 6.4×10−3counts/neutron, respectively, for 2.5 MeV‐DD and 14 MeV‐DT neutrons. The apparently higher efficiency observed for the 14 MeV neutrons is attributed to neutron‐induced background gamma radiation.