Fast Fission Chamber Amplifier‐Discriminator System
作者:
R. W. Ingle,
F. E. Gillespie,
L. W. Weston,
期刊:
Review of Scientific Instruments
(AIP Available online 1970)
卷期:
Volume 41,
issue 11
页码: 1539-1543
ISSN:0034-6748
年代: 1970
DOI:10.1063/1.1684337
出版商: AIP
数据来源: AIP
摘要:
To measure the fission cross section of233U as a function of neutron energy, instrumentation with fast pulse rise‐time and short resolving time characteristics is required. The efficient detection of fission‐fragment pulses in the presence of a high alpha background requires these characteristics. A fission detector with an electron collection time of 25 nsec and containing 1 g of233U, and an amplifier with a risetime of 7 nsec and an over‐all gain of 1000 were developed for this measurement. This system, detecting the fast rising current pulse from a parallel plate ionization chamber, is an integral part of the instrumentation used by a group at ORNL engaged in the measurement of the neutron fission and capture cross sections as a function of neutron energy. With this system, the fission cross section of233U was successfully measured over a neutron energy range from 0.4 to 2000 eV.