Modulation Pickup in Nuclear Magnetic Resonance rf Spectrometers

 

作者: G. B. Robinson,   F. E. Geiger,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1958)
卷期: Volume 29, issue 8  

页码: 730-731

 

ISSN:0034-6748

 

年代: 1958

 

DOI:10.1063/1.1716310

 

出版商: AIP

 

数据来源: AIP

 

摘要:

Radio‐frequency oscillators used as detectors of nuclear paramagnetic resonance absorption employ an auxiliary source at audio frequency to modulate the static magnetic field. A small component of voltage at this frequency is usually induced into the tank coil of the oscillator. If this voltage is permitted to appear at the grid of the oscillator, mixing with the rf voltage will occur. This produces a false signal indistinguishable from the resonance signal. A simple LC filter put between tank circuit and oscillator tube is found to be extremely effective in removing false resonances. The response of the oscillator to signals generated by a Watkins calibrator is measured in the range from 5 to 25 Mc. Results of these and ``Q'' measurements show the oscillator sensitivity to be unimpaired by the addition of the LC network.

 