首页   按字顺浏览 期刊浏览 卷期浏览 Calibration data for the Telex 1470‐A audiometric earphones
Calibration data for the Telex 1470‐A audiometric earphones

 

作者: Paul L. Michael,   Gordon R. Bienvenue,  

 

期刊: The Journal of the Acoustical Society of America  (AIP Available online 1980)
卷期: Volume 67, issue 5  

页码: 1812-1815

 

ISSN:0001-4966

 

年代: 1980

 

DOI:10.1121/1.384259

 

出版商: Acoustical Society of America

 

数据来源: AIP

 

摘要:

Standard reference threshold levels for the Telex 1470‐A audiometric earphones have been determined. The psychoacoustic evaluation of these earphones was completed by loudness balancing at threshold. The earphones were balanced to comparison standard earphones calibrated at the National Bureau of Standards. Coupler calibration was accomplished using the American National Standard Method. Objective (coupler) and subjective (psychoacoustic) data were combined and the resulting reference threshold sound pressure level data are reported herein. These data were compared to previous data for the 1470‐type earphone. While good agreement with previous data was noted for low and mid frequencies, sizeable differences did occur at high frequencies. Possible sources of these discrepancies are discussed.

 

点击下载:  PDF (495KB)



返 回