首页   按字顺浏览 期刊浏览 卷期浏览 Flow-Injection Chemiluminescence Sensor for the Determination of Free Chlorine in Tap W...
Flow-Injection Chemiluminescence Sensor for the Determination of Free Chlorine in Tap Water

 

作者: Wei Qin,   Zhujun Zhang,   Shuna Liu,  

 

期刊: Analytical Letters  (Taylor Available online 1997)
卷期: Volume 30, issue 1  

页码: 11-19

 

ISSN:0003-2719

 

年代: 1997

 

DOI:10.1080/00032719708002286

 

出版商: Taylor & Francis Group

 

关键词: chemical sensor;chemiluminescence;flow-injection;free chlorine

 

数据来源: Taylor

 

摘要:

A novel chemiluminescence(CL) sensor combined with flow-injection analysis has been developed for determining free chlorine in tap water. The analytical reagent luminol was immobilized on an anion exchange resin column. While a volumn of sodium hydroxide passed through the column, luminol was eluted from the resin in alkaline aqueous solution and then mixed with a sample stream to react and produce CL. The CL emission intensity was correlated with the standard ClO−concentration in the range 1×10−8to 4×10−5g ml−1, and the detection limit was 8×10−9g ml−1ClO−. Interfering metal ions present in water were effectively separated by a pre-column cation exchanger. A complete analysis, including sampling and washing, could be performed in 1 min with a relative standard deviation of less than 5%. The sensor was stable for over 200 times and has been applied successfully to the determination of ClO−in tap water.

 

点击下载:  PDF (288KB)



返 回