Frequency Scan of a Quadrupole Mass Analyzer in the Third Stability Region for Protein Analysis
作者:
Z. X. Nie,
C.‐W. Lin,
W.‐P. Peng,
Y. T. Lee,
H.‐C. Chang,
期刊:
Journal of the Chinese Chemical Society
(WILEY Available online 2006)
卷期:
Volume 53,
issue 1
页码: 47-52
ISSN:0009-4536
年代: 2006
DOI:10.1002/jccs.200600006
出版商: WILEY‐VCH Verlag
关键词: Quadrupole mass analyzer;Frequency scan;Proteins
数据来源: WILEY
摘要:
AbstractA linear quadrupole mass spectrometer operating in the third stability region [(a, q) ≈ (3, 3)] has been developed and utilized to obtain the mass spectra of multiply charged protein ions generated by electro‐spray ionization. The spectra were acquired by scanning the frequency of the quadrupole with a home‐built wide‐band power amplifier from 400 to 100 kHz at constant voltage amplitudes in the range of 100 V. A mass resolution approachingR1/2= 200 was achieved for doubly charged gramicidinSions atm/z570.5. Further improvement of the resolution is anticipated when a higher‐voltage power amplifier is