Gas chromatographic method for simulated distillation up to a boiling point of 750°C using temperature‐programmed injection and high temperature fused silica wide‐bore columns
作者:
J. Curvers,
P. van den Engel,
期刊:
Journal of High Resolution Chromatography
(WILEY Available online 1989)
卷期:
Volume 12,
issue 1
页码: 16-22
ISSN:0935-6304
年代: 1989
DOI:10.1002/jhrc.1240120106
出版商: Wiley Subscription Services, Inc., A Wiley Company
关键词: Gas chromatography;Wide bore capillary columns;Temperature programmed injection;Simulated distillation
数据来源: WILEY
摘要:
AbstractIt is demonstrated that linear injection characteristics are obtained for a wide boiling point range sample using a temperature‐programmed injector in combination with wide‐bore fused silica columns. The applicability of the described combination for high temperature simulated distillation is described. The method, using external standardization, gives accurate and repeatable results for different types of samples in the boiling range between 50 and 750°C. The lifetime of the fused silica wide‐bore columns was found to be acceptable,viz.over 80 temperature‐programmed cycles between ambient and 430°C. Some comments are made on the accuracy of boiling points for norma
点击下载:
PDF
(715KB)
返 回