首页   按字顺浏览 期刊浏览 卷期浏览 Experimental determination of the need for multitaper collimators for neutron radiother...
Experimental determination of the need for multitaper collimators for neutron radiotherapy

 

作者: K. M. Jones,   C. A. Kelsey,  

 

期刊: Medical Physics  (WILEY Available online 2003)
卷期: Volume 1, issue 4  

页码: 215-218

 

ISSN:0094-2405

 

年代: 2003

 

DOI:10.1118/1.1637298

 

出版商: American Association of Physicists in Medicine

 

数据来源: WILEY

 

摘要:

A 52.5‐cm‐long steel collimator for 15‐MeV neutron therapy beams has been evaluated in terms of the distribution of the collimator‐scattered neutrons in the field and penumbra, the distribution in the collimator of the neutron scattering sites, and the energy of the neutrons scattered to one position in the penumbra, in order to assess the need for multiple tapers on such collimators. The associated particle technique was used to eliminate background neutrons from the measurement. We conclude that because collimator‐scattered neutrons constitute only a small part of the neutron flux in the field and most of the scattered neutrons have energies greater than 12 MeV, double‐taper collimators are not necessary.

 