Improved energy resolution for scintillator based gamma-ray detectors obtained using a CsI(Tl)/Si-PINdetector
作者:
B. E. Patt,
J. S. Iwanczyk,
期刊:
Review of Scientific Instruments
(AIP Available online 1997)
卷期:
Volume 68,
issue 10
页码: 3926-3927
ISSN:0034-6748
年代: 1997
DOI:10.1063/1.1148051
出版商: AIP
数据来源: AIP
摘要:
This note reports improved energy resolution measured using a gamma-ray detector consisting of a 2 mm×2 mm×4 mm long CsI(Tl) scintillator coupled to a 2 mm×2 mm photodiode. Energy resolution of 7.5&percent; full width at half-maximum (FWHM) was obtained for the 122 keV photopeak of Co-57. Energy resolution of 35&percent; FWHM was obtained for the 5.9 keV photopeak of Fe-55. ©1997 American Institute of Physics.
点击下载:
PDF
(35KB)
返 回