首页   按字顺浏览 期刊浏览 卷期浏览 Background correction in electron-ion coincidence experiments using a self-optimizing, ...
Background correction in electron-ion coincidence experiments using a self-optimizing, pseudorandom count generator

 

作者: Ivan Powis,   Peter Downie,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1998)
卷期: Volume 69, issue 9  

页码: 3142-3145

 

ISSN:0034-6748

 

年代: 1998

 

DOI:10.1063/1.1149074

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A technique is described whereby an estimate of the false coincidence signal, suitable for background correction of data acquired in a coincidence experiment, is obtained by using a pseudorandom pulser to generate a stream of “false” start events. The statistical properties of this simulated source are adjusted to mimic those of the real source of electron start events. False ion coincidences with the simulated starts are measured concurrently with the real coincidence signal, with the mean count rate of the pseudorandom pulse source automatically tracking that of the true electron start events. In this manner any long term instrumental drifts during the course of an extended experimental measurement will similarly affect both the real and simulated coincidence data. Subtraction of the simulated background of false coincidences from the real coincidence data then yields an improved estimate of the true coincidence signal. ©1998 American Institute of Physics.

 

点击下载:  PDF (71KB)



返 回