Differences Between ac and dc Determinations of Germanium Thermometer Resistances
作者:
C. A. Swenson,
P. C. F. Wolfendale,
期刊:
Review of Scientific Instruments
(AIP Available online 1973)
卷期:
Volume 44,
issue 3
页码: 339-341
ISSN:0034-6748
年代: 1973
DOI:10.1063/1.1686127
出版商: AIP
数据来源: AIP
摘要:
The resistances of several germanium thermometers are found to differ for ac and dc measurements. At 300 K, these differences are frequency independent from roughly 5 to 103Hz. The ac resistances always are smaller, and the approximate temperature differences (Tdc−Tac) vary from 100 mk at 77 K to 0.2 mK below 17 K.