Accuracy of a mixed venous saturation catheter during acutely induced changes in hematocrit in humans

 

作者: ELISABETH VAN WOERKENS,   ADRIANUS TROUWBORST,   ROB TENBRINCK,  

 

期刊: Critical Care Medicine  (OVID Available online 1991)
卷期: Volume 19, issue 8  

页码: 1025-1029

 

ISSN:0090-3493

 

年代: 1991

 

出版商: OVID

 

关键词: hemodilution;pulmonary wedge pressure;fiber optics;hematocrit;cardiac output;catheterization;hemoglobin;laparotomy;blood gas analysis

 

数据来源: OVID

 

摘要:

ObjectiveTo determine the accuracy of in vivo mixed venous hemoglobin saturation (S&OV0456;o2) measurements with a fiber optic thermodilution catheter during acute changes in hematocrit.DesignComparison of fiberoptic in vivo S&OV0456;o2values with in vitro S&OV0456;o2values obtained with a multiwavelength spectrophotometer.SettingOperating room in a university hospital.PatientsSix consecutive patients who are Jehovah's Witnesses.Measurements and Main ResultsBefore and after each step of hypervolemic hemodilution and after every 500 mL of blood loss, blood gases were analyzed and hemodynamic, hemoglobin, hematocrit, and in vitro and in vivo S&OV0456;o2measurements were made. Hematocrit values were measured in the range of 40% to 18%. Plotting all in vivo S&OV0456;o2values (n = 74) against the in vitro S&OV0456;o2measurements obtained during the entire study period gives r2= .86. The accuracy of in vivo S&OV0456;o2measurements was not affected by changes in hematocrit or cardiac output. The S&OV0456;o2catheter value at the beginning of the study differed from the in vitro S&OV0456;o2value by −0.86 ± 2.56% and at the end of the study period of 8 to 10 hrs by 0.71 ± 3.04%.ConclusionsThe accuracy of the studied fiberoptic continuous measuring S&OV0456;o2system was not affected by changes in hematocrit or cardiac output. No significant drift in the in vitro S&OV0456;o2measurements was observed.

 

点击下载:  PDF (403KB)



返 回