首页   按字顺浏览 期刊浏览 卷期浏览 Simpler and More Sensitive Immune Complex Transfer Enzyme Immunoassay for Anti-Htlv-I I...
Simpler and More Sensitive Immune Complex Transfer Enzyme Immunoassay for Anti-Htlv-I Igg Using Modified Polystyrene Beads, Microplates and Fluororeader

 

作者: Setsuko Ishikawa,   Seiichi Hashida,   Hidetaka Nakamoto,   Satoshi Tanaka,   Massaki Kojima,   Eiji Ishikawa,  

 

期刊: Analytical Letters  (Taylor Available online 1995)
卷期: Volume 28, issue 9  

页码: 1611-1618

 

ISSN:0003-2719

 

年代: 1995

 

DOI:10.1080/00032719508002768

 

出版商: Taylor & Francis Group

 

关键词: Enzyme immunoassay;Antibody IgG;HTLV-I;Carryover;Microplate;Fluororeader

 

数据来源: Taylor

 

摘要:

In the previous immune complex transfer enzyme immunoassay for anti-HTLV-I IgG, the transfers of polystyrene beads in and out of test tube were handled with tweezers, and the bound β-D-galactosidase activity was measured with a fluorometer. The use of tweezers was considered the primary cause of false-positivity by carryover. Furthermore, the testing of many samples using a tweezer as a means of transfer was difficult. In the present immune complex transfer enzyme immunoassay, polystyrene beads were attached to plates through cylindrical bars and were used in microplate wells. Therefore, no tweezers were required. The bound β-D-galactosidase activity was measured with a fluororeader. This allowed the elimination of false-positivity due to carryover and made it easier to test many samples with higher sensitivity and reliability.

 

点击下载:  PDF (245KB)



返 回