Use of the Four‐Inch Liquid Hydrogen Bubble Chamber as a Fast‐Neutron Spectrometer

 

作者: Harold E. Adelson,   Hoyt A. Bostick,   Burton J. Moyer,   Charles N. Waddell,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1960)
卷期: Volume 31, issue 1  

页码: 1-10

 

ISSN:0034-6748

 

年代: 1960

 

DOI:10.1063/1.1716782

 

出版商: AIP

 

数据来源: AIP

 

摘要:

The 4‐in. hydrogen bubble chamber has been used as a neutron spectrometer in the 5‐ to 30‐Mev region. A fast data‐reduction system for the analysis of the recoil‐proton tracks was developed, utilizing a commercial electronic coordinate‐measuring device, the IBM punch‐card system, and the IBM 650 digital computer. The resolution of the entire spectrometer system was determined by measuring the monoergic (14.1‐Mev) neutron spectrum from the t(d, n)He4reaction with different neutron collimators. The full width at half maximum of the measured peak was 10% of the peak energy.

 