Fatal cardiac tamponade and other mechanical complications of central venous catheters
作者:
T. E. B. Dane,
E. G. King,
期刊:
British Journal of Surgery
(WILEY Available online 1975)
卷期:
Volume 62,
issue 1
页码: 6-10
ISSN:0007-1323
年代: 1975
DOI:10.1002/bjs.1800620103
出版商: John Wiley&Sons, Ltd.
数据来源: WILEY
摘要:
AbstractA fatal cardiac tamponade caused by an intact polyethylene central venous catheter perforating the wall of the right ventricle is reported. A retrospective review of 105 central venous catheters used in an intensive care unit revealed a high incidence of catheter malposition and a 20 per cent mechanical complication rate. The possible causes of these malpositions and complications are discussed. Guidelines for the safer use of central venous catheters are presented.