A Sensitive Gas Chromatograph Detector for Nitrosamines
作者:
Yin Fang,
Ding Jia-hua,
Liu Shi-li,
期刊:
Analytical Letters
(Taylor Available online 1981)
卷期:
Volume 14,
issue 12
页码: 977-983
ISSN:0003-2719
年代: 1981
DOI:10.1080/00032718108055494
出版商: Taylor & Francis Group
关键词: N-nitrosamines;Food;Gas Chromatography
数据来源: Taylor
摘要:
A GC method using a flame thermionic detector (FTD) with a RbSO4source to determine nitrosamines is described. The method is sensitive enough to be used for a preliminary screening of foodstuffs for the presence of volatile nitrosamines. The lowest detection limit is 0.1 ng for DMNA, DENA, or DPNA, with a linear range of 0.1 − 100 ng determinable. The method has a favorable stability; the factors affecting its sensitivity, e.g. the H2flow rate, column temperature and the solvent used, have been studied.
点击下载:
PDF (184KB)
返 回