Rapid and Sensitive Spectrophotometry. II. A Stopped‐Flow Attachment for a Stabilized Quartz Spectrophotometer
作者:
Britton Chance,
Victor Legallais,
期刊:
Review of Scientific Instruments
(AIP Available online 1951)
卷期:
Volume 22,
issue 8
页码: 627-634
ISSN:0034-6748
年代: 1951
DOI:10.1063/1.1746020
出版商: AIP
数据来源: AIP
摘要:
The stabilization of the light sources and the photocell circuits of a quartz spectrophotometer and the design of suitable stopped flow apparatuses are described. The latter have been designed for 1.6‐ and 10‐mm optical paths and have been shown to give rapid mixing (<0.1 sec). Errors of optical density measurement are diminished by rectangular observation tubes, and errors of extinction coefficients are minimized by a narrow spectral interval (less than 2 m&mgr; from 230–600 m&mgr;). The error of optical density caused by the photoelectric circuits is about 1×10−4for a response time (10–90 percent) of 1 sec. The application of this apparatus to the measurement of rapid chemical reactions in the visible and ultraviolet regions is described.