首页   按字顺浏览 期刊浏览 卷期浏览 Novel method of compensating shunts for measuring kiloampere pulsed currents
Novel method of compensating shunts for measuring kiloampere pulsed currents

 

作者: U. K. Chatterjee,   N. S. Shikarkhane,   U. Nundy,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1981)
卷期: Volume 52, issue 4  

页码: 618-620

 

ISSN:0034-6748

 

年代: 1981

 

DOI:10.1063/1.1136652

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A simple method of compensating an inductive strip shunt is to use another shunt in series with it, with equal inductance but lesser resistance. When the voltages induced in these shunts are subtracted in a compensating circuit, the compensated shunt acts like a pure resistance. A compensated shunt of thin tantalum foil, having an effective resistance of 200 mW, was made.

 

点击下载:  PDF (185KB)