Shock Tube Studies Using Image Orthicon Spectroscopy
作者:
E. W. Hewitt,
J. F. Bott,
T. A. Jacobs,
期刊:
Review of Scientific Instruments
(AIP Available online 1970)
卷期:
Volume 41,
issue 10
页码: 1416-1418
ISSN:0034-6748
年代: 1970
DOI:10.1063/1.1684296
出版商: AIP
数据来源: AIP
摘要:
An image orthicon detection and recording system has been adapted for making spectroscopic measurements in shock tube emission studies. The system provides greater sensitivity than film and permits electronic control of both the timing and duration of the exposure. The linear response with light intensity and immediate recording of output on an oscilloscope trace makes the system a convenient diagnostic tool. A description of the system as well as quantitative spectroscopic data obtained with it is presented.