首页   按字顺浏览 期刊浏览 卷期浏览 Determination of Ammonium by Using a Flow System with a Diffusion Cell Coupled to an Am...
Determination of Ammonium by Using a Flow System with a Diffusion Cell Coupled to an Ammonium Ion-Selective Sensor

 

作者: Renmin Liu,   Bianting Sun,  

 

期刊: Analytical Letters  (Taylor Available online 1997)
卷期: Volume 30, issue 7  

页码: 1255-1265

 

ISSN:0003-2719

 

年代: 1997

 

DOI:10.1080/00032719708007214

 

出版商: Taylor & Francis Group

 

关键词: Ammonium determination;Potentiometry;Ammonium electrode;Gas diffusion

 

数据来源: Taylor

 

摘要:

This paper describes a flow potentiometric method for determination of ammonium at low levels. A new diffusion cell coupled to an ammonium ion electrode was used as the potentiometric detector. This method eliminates interference to the ammonium ion electrode. When the stopped-flow method is employed for the acceptor stream, ammonium ions can be accumulated in the acceptor solution in the diffusion cell, giving high selectivity and sensitivity. The lower linear value of the proposed method can reach 1 × 10−6mol L−1with an accumulation time of 5 min. The method is simple, sensitive and suitable for online determination of ammonium in water samples. It has been applied to the determination of ammonium in rain water and river water, and recoveries of 95 – 104% were obtained. The relative standard deviation of seven replicate analyses for all samples was 1.1 – 2.7%.

 

点击下载:  PDF (336KB)



返 回