High Resolution Gel Permeation Chromatography-Using Recycle

 

作者: KarlJ. Bombaugh,   RobertF. Levangie,  

 

期刊: Separation Science  (Taylor Available online 1970)
卷期: Volume 5, issue 6  

页码: 751-763

 

ISSN:0037-2366

 

年代: 1970

 

DOI:10.1080/00372367008055537

 

出版商: Taylor & Francis Group

 

数据来源: Taylor

 

摘要:

High resolution in gel permeation chromatography has been accomplished by use of long columns, since commercial rigid gels afford relatively fixed capacity ratios and since plate numbers of equivalent columns are additive. To overcome the high cost and high pressure requirement of long columns often required to resolve discrete species, high resolution gel permeation chromatography may be attained in commercial GPC equipment by recycle operation through the reciprocating pump, GPC column, and detector. However, peak width increases with the number of cycles (μ). Since the contained volume of the closed recycle system is constant, asvincreases the peak width (W) of the distribution will eventually exceed the volume of the system and peak overlap will occur. This presentation considers the increase inWand provides a method of “flush and draw off” to prevent sample overlap. Analytical and preparative scale separations were investigated, using both small and macromolecules. The effects of sample load and flow rate on-resolution with recycle operation were investigated.

 