The Effect of Abutment Angulation on Stress Transfer for an Implant
作者:
Nancy L. Clelland,
Amos Gilat,
期刊:
Journal of Prosthodontics
(WILEY Available online 1992)
卷期:
Volume 1,
issue 1
页码: 24-28
ISSN:1059-941X
年代: 1992
DOI:10.1111/j.1532-849X.1992.tb00422.x
出版商: Blackwell Publishing Ltd
关键词: stress transfer;photoelasticity;angled abutments;strain gauges
数据来源: WILEY
摘要:
PurposeThis investigation compared the stress production characteristics of five abutment angulations for a specific implant system.Materials and MethodsPhotoelastic resin was cast directly to a 3.75 × 10‐mm Branemark fixture (Nobelpharma USA, Inc, Chicago, IL) in a 50 × 70 × 13‐mm mold. A strain gauge rosette was also incorporated in the resin to allow precise determination of normal stresses at a specific point. Each 4‐mm abutment (15°, 25°, and 35° from Implant Innovations (West Palm Beach, FL) and 0° and 30° from Nobelpharma) was assembled on the fixture, subjected to 178N load, and viewed with a circular polariscope. Observed fringes were photographed and strain indicator readings were recorded.ResultsMean observed fringe order and mean principal stress and strain at the location of the rosette were determined.ConclusionsAt the location of the rosette, all five of the abutments produced principal strains (compressive and tensile) within the physiological zone for bone. The rosette was located approximately 4 mm away from the fixture. Higher stresses and strains can be expected in regions closer t