Low‐Temperature Adiabatic Calorimeter with Automatic Shield Control

 

作者: W. J. Cooper,   R. H. Forsyth,   J. F. Masi,   R. E. Pabst,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1959)
卷期: Volume 30, issue 7  

页码: 557-559

 

ISSN:0034-6748

 

年代: 1959

 

DOI:10.1063/1.1716680

 

出版商: AIP

 

数据来源: AIP

 

摘要:

The design and operation of an adiabatic low‐temperature calorimeter employing a somewhat different design of automatic shield control is described. Thermocouple difference is sensed by a dc &mgr;v amplifier, which is followed in the circuit by a recorder‐controller and a Leeds & Northrup Series 60 (current adjusting type) control unit. The output of the latter is used to control the voltage across the shield heaters through a Westinghouse ``Magamp'' saturable reactor. This design has been used successfully with a conventional low‐temperature adiabatic calorimeter (50‐ml sample size). In most cases the recorder showed no variation from zero (±0.002°) during heating and equilibrating periods, while momentary disturbance at ``on'' and ``off'' was less than 0.05°.

 