Automatic Calorimetry in the 3–30 K Range. The Specific Heat of Copper

 

作者: Douglas L. Martin,   L. L. T. Bradley,   W. J. Cazemier,   Robert L. Snowdon,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1973)
卷期: Volume 44, issue 6  

页码: 675-684

 

ISSN:0034-6748

 

年代: 1973

 

DOI:10.1063/1.1686220

 

出版商: AIP

 

数据来源: AIP

 

摘要:

Automatic data acquisition is by means of a slightly modified version of a hard‐wired system described previously (Martin and Snowdon, 1970) combined with an on‐line minicomputer which controls the calorimeter heating rate and computes temperature and specific heat. The new cryostat, calorimeter assembly, and adiabatic shield control are briefly described. Experiments to verify the accuracy of germanium thermometry with the automatic ac bridge are detailed. Operation of the system is illustrated by results on pure copper which show a standard deviation of 0.3% (mostly due to temperature scale defects rather than inherent lack of precision) and agree well with the Copper Reference Equation below 20 K but above this temperature agree more closely with the results of Cetas, Tilford, and Swenson[Phys. Rev. 174, 835 (1968)].

 