首页   按字顺浏览 期刊浏览 卷期浏览 Air Alpha Proportional Counter Insensitive to Atmospheric Humidity
Air Alpha Proportional Counter Insensitive to Atmospheric Humidity

 

作者: A. M. R. Ferrari,   C. J. Borkowski,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1962)
卷期: Volume 33, issue 10  

页码: 1047-1050

 

ISSN:0034-6748

 

年代: 1962

 

DOI:10.1063/1.1717678

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A conventional alpha proportional counter which uses air as the counter gas is sensitive to high relative humidity and generates spurious pulses that cannot be distinguished from actual alpha pulses. It has been found possible to operate such a counter satisfactorily at high relative humidity by passing a small current (∼15 mA) through the center wire. In this manner the center wire is heated and the relative humidity of the surrounding sheath of air is reduced sufficiently so that operation of the counter at high relative humidity is comparable to operation with dry air. Two different mechanisms are proposed for the formation of spurious pulses in such a counter at high relative humidity.

 

点击下载:  PDF (307KB)



返 回