首页   按字顺浏览 期刊浏览 卷期浏览 Flow Injection Continuous Liquid-Liquid Extraction Using a Hollow Fiber Membrane Separa...
Flow Injection Continuous Liquid-Liquid Extraction Using a Hollow Fiber Membrane Separator for Determination of Copper in Water

 

作者: Pablo Richter,   M. Inés Toral,   Roberto Manríquez,  

 

期刊: Analytical Letters  (Taylor Available online 1999)
卷期: Volume 32, issue 3  

页码: 601-612

 

ISSN:0003-2719

 

年代: 1999

 

DOI:10.1080/00032719908542842

 

出版商: Taylor & Francis Group

 

关键词: Flow injection;continuous liquid-liquid extraction;hollow fiber;copper, water

 

数据来源: Taylor

 

摘要:

A simple flow injection photometric method for the determination of copper was developed based on the formation and liquid-liquid extraction of the analyte-bathocuproine-perchlorate ternary complex. The analytical reaction of complex formation is integrated with the continuous liquid-liquid extraction procedure. A hollow fiber-type membrane separator was specially constructed for the determination. The analytical properties like sensitivity, precision, accuracy and rapidity were evaluated. The sensitivity of the method depends on the injection volume. By using an injection volume of 0.6 ml, the detection limit (3σ) found was 15 ng ml−1copper, with a determination range of 48 - 3000 ng/ml. The repeatability of the method, expressed as the relative standard deviation, was in all instances less than 3.0%. The sampling rate was of 20 h−1. The accuracy of the method was tested with certified reference materials, and the mean error was calculated to be less than 5%. The method was successfully applied to the determination of copper in river water.

 

点击下载:  PDF (377KB)



返 回