作者: G F Watts,   J Quiney,   R W Morris,   P Thaker,   S Craddock,   K M Shaw,  

 

期刊: Practical Diabetes International  (WILEY Available online 1988)
卷期: Volume 5, issue 1  

页码: 33-35

 

ISSN:1357-8170

 

年代: 1988

 

DOI:10.1002/pdi.1960050112

 

出版商: John Wiley&Sons, Ltd.

 

数据来源: WILEY

 

摘要:

AbstractA quality control filter‐paper method was applied to the assessment of two glucose enzyme reagent strips — Type A and Type B — in the home environment. Forty insulin‐dependent diabetics, experienced in home blood glucose monitoring, took part in a randomised, cross‐over trial. Marked variation was observed in the accuracy of readings with both strips over a wide range of glucose concentrations. Overall, Type B was found to read consistently higher than the reference filter‐paper method, whereas Type A read consistently higher for reference glucose concentrations<10 mmol/l, but lower for concentrations ≥10 mmol/l. In particular, both strips were equally poor at detecting filter‐paper concentrations in the hypoglycaemic range. Although both strips have been modified since the study was initiated, the principle of quality control is equally applicable to all other currently used and visually read strips. Recommendations are provided for the quality assessment of all reagent strips employed by patients for home blood glu

 