Effects of frequency and magnetic field scaling on the superconducting electron cyclotron resonance ion source at MSU‐NSCL
作者:
S. Gammino,
G. Ciavola,
R. Harkewicz,
K. Harrison,
A. Srivastava,
P. Briand,
期刊:
Review of Scientific Instruments
(AIP Available online 1996)
卷期:
Volume 67,
issue 12
页码: 4109-4113
ISSN:0034-6748
年代: 1996
DOI:10.1063/1.1147556
出版商: AIP
数据来源: AIP
摘要:
A series of systematic tests have been carried out on the superconducting electron cyclotron resonance ion source SCECRIS at the Michigan State University National Superconducting Cyclotron Laboratory in order to better understand the role of different parameters which determine the performance of these sources. A relevant part of these tests has been focused on the role of magnetic field and microwave frequency, with the goal of demonstrating the capability of low frequency electron cyclotron resonance ion sources to produce high ion charge states, in spite of their relative low plasma density. This goal can be achieved using a very high magnetic confining field, according to the concept of a high B mode. A comparison of the source performance at 2.45 and 6.4 GHz for different levels of magnetic field is described in this article. ©1996 American Institute of Physics.