首页   按字顺浏览 期刊浏览 卷期浏览 Tensiometer for free standing smectic films
Tensiometer for free standing smectic films

 

作者: Matthias Eberhardt,   Robert B. Meyer,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1996)
卷期: Volume 67, issue 8  

页码: 2846-2851

 

ISSN:0034-6748

 

年代: 1996

 

DOI:10.1063/1.1147116

 

出版商: AIP

 

数据来源: AIP

 

摘要:

We present the design of a machine which allows us to measure the surface tension of a free standing smectic film with an accuracy of 0.1 mN/m absolute and 0.01 mN/m relative. Our design makes it possible to measure the surface tension over a wide range of temperatures with minimal drift in the detector itself. The response time of our detector can be as small as 0.01 s which allows us to study dynamic processes involving the interaction between the film and the meniscus around the frame that supports the film. Our design is based on a rigid frame with a movable side suspended as a pendulum. The surface tension force on the movable side is opposed by a feedback system to stabilize the position of the movable side. The stabilizing force then is a measure of the surface tension. An optical system detects deflection of the movable side due to the surface tension force, and applies a restoring force by generating electric current in a coil which applies force to a permanent magnet attached to the pendulum arm. The feedback control includes an integrator stage to achieve finite force with negligible error in position. The device can be calibrated with a simple lever and weight scheme to simulate a surface tension force. The design elements that determine the stability and accuracy of the system are discussed. ©1996 American Institute of Physics.

 

点击下载:  PDF (132KB)