Theoretical and experimental sensitivities of ESR spectrometers using maser techniques
作者:
J. C. Mollier,
J. Hardin,
J. Uebersfeld,
期刊:
Review of Scientific Instruments
(AIP Available online 1973)
卷期:
Volume 44,
issue 12
页码: 1763-1771
ISSN:0034-6748
年代: 1973
DOI:10.1063/1.1686050
出版商: AIP
数据来源: AIP
摘要:
We have considered theoretically and experimentally a paramagnetic resonance spectrometer, whose cavity contains both the sample to be studied and a beam of ammonia molecules capable of stimulated emission. Different modes of operation are investigated: classical, ``Q‐multiplier'' and marginal oscillator spectrometer. In these two latter cases, the ultimate sensitivity is determined only by the thermal noise in the cavity. It is shown to be the theoretical sensitivity of a spectrometer using a perfect receiver. This sensitivity can be achieved if the stored energy does not exceed an upper limit, depending on properties of the active molecules‐typically 5 × 10−18J for NH3.