Respiratory Failure Treated in the Hyperbaric Chamber with the Membrane OxygenatorIII. Studies of Three Membranes at Various Flow Rates and Ambient Pressures

 

作者: Barbara Lipton,   Avron Weinreich,   Too Brondum,   Julius Jacobson,  

 

期刊: Anesthesiology  (OVID Available online 1971)
卷期: Volume 34, issue 5  

页码: 421-426

 

ISSN:0003-3022

 

年代: 1971

 

出版商: OVID

 

关键词: Respiratory failure;Oxygen toxicity;Hyperbaric chamber;Membrane oxygenator

 

数据来源: OVID

 

摘要:

Increasing partial pressure of oxygen by hyperbaric oxygenation improves the efficiency of membrane oxygenators sufficient to provide almost total respiratory support with membranes of practical sizes. The oxygenating capacities of the Teflon-Peirce, Landé-Edwards and G.E.-Peirce membrane oxygenators were tested using surface areas of 1 to 3 m2, flow rates of 500 to 2,000 ml/ min, and ambient pressures from 1 to 3 ATA (atmospheres absolute) in the hyperbaric chamber. For every 3 m2-membrane tested at 2 ATA or more the amount of oxygen in the blood leaving the oxygenator approached its maximum oxygencarrying capacity. The Teflon-Peirce membrane decreased in efficiency with increasing flow rates and is unsuitable for clinical use. The Landé-Edwards (3 m2) and the G.E.-Peirce (2 m2) at flow rates of 2,000 ml at 2 ATA were capable of transferring oxygen at the rates of 200 and 240 ml/min, respectively. It was not necessary to pressurize above 2 ATA to achieve efficient use of the membrane oxygenators tested.

 