Three‐Way Comparison of Retinoscopy, Subjective, and Dioptron Nova Refractive Findings
作者:
THEODORE GROSVENOR,
DAVID PERRIGIN,
JUDITH PERRIGIN,
期刊:
American Journal of Optometry and Physiological Optics
(OVID Available online 1985)
卷期:
Volume 62,
issue 1
页码: 63-65
ISSN:0093-7002
年代: 1985
出版商: OVID
关键词: retinoscopy;Dioptron Nova refraction;refractive error
数据来源: OVID
摘要:
ABSTRACTOn the basis of routine clinical findings taken on a group of 100 patients serving as subjects in the Houston Myopia Control study, a three-way comparison was made between retinoscopy, subjective, and Dioptron Nova refractive findings. Retinoscopy and subjective findings compared very favorably, mean differences being –0.01 D for spherical equivalent power and –0.05 D for cylinder power. Dioptron and retinoscopy findings differed by –0.30 D for spherical equivalent power and –0.13 D for cylinder power, whereas Dioptron and subjective findings differed by –0.33 D for spherical equivalent power and –0.18 D for cylinder power.