首页   按字顺浏览 期刊浏览 卷期浏览 Accuracy of blood transit time estimates using temporal filtering in digital radiology
Accuracy of blood transit time estimates using temporal filtering in digital radiology

 

作者: D. R. Wyman,   B. C. Wilson,  

 

期刊: Medical Physics  (WILEY Available online 1998)
卷期: Volume 13, issue 1  

页码: 8-12

 

ISSN:0094-2405

 

年代: 1998

 

DOI:10.1118/1.595928

 

出版商: American Association of Physicists in Medicine

 

数据来源: WILEY

 

摘要:

The application of recursive filtering in digital radiology can yield incorrect blood flow estimates if blood flow is calculated using a transit time, defined as the difference in times to peak contrast over the length of a vessel segment. The transit time error is calculated here, from computer simulations, as a function of filter design, image noise and contrast medium dispersion over the vessel segment. An error reduction procedure based on prefiltering is suggested and tested by simulation. Results indicate that the transit time error can be substantial and the reduction procedure effective, depending on the filter design and system noise level.

 

点击下载:  PDF (463KB)



返 回